Thermo Scientific Dionex ICS-5000 Operator's Manual page 304

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 Operator's Manual
No chip card in the reader for the column x.
This error occurs if a column ID chip card is not installed for the column
identified in the message.
To troubleshoot:
Install a column ID chip card for the column (see
not want to store information about this column, follow the steps below to
disable this option.
1. In the Chromeleon 7 Instrument Configuration Manager or the
Chromeleon 6.8 Server Configuration program, right-click the TC and
select Properties on the context menu.
2. Select the Components tab in the Properties dialog box. Clear the check
mark beside the column name and click OK.
No response from x for x seconds.
-or-
No response from x for x seconds. Device disconnected.
This error occurs if there is a communication failure between the TC and the
PC on which Chromeleon is installed.
To troubleshoot:
1. Check the USB connections.
2. Check the connection from the TC to the power supply.
3. If you cannot identify the cause of the error, contact Technical Support for
Dionex products for assistance.
288
Section
9.30). Or, if you do
Doc. 065342-06 12/11