Press the
turn off the pump.
3. Turn off the main power switch on the rear panel of the pump (see
Figure
4. Disconnect the main power cord from both its source and from the rear
panel of the pump.
HIGH VOLTAGE—Disconnect the main power cord from its source and
also from the rear panel of the pump.
HAUTE TENSION—Débranchez le cordon d'alimentation principal de sa
source et du panneau arrière du pump.
HOCHSPANNUNG—Ziehen Sie das Netzkabel aus der Steckdose und
der Netzbuchse auf der Rückseite der Pumpe.
5. The fuse cartridge is located next to the
main power switch (see
Use a small screwdriver to remove the
fuse cartridge.
6. Replace the two fuses with new 2 A IEC
60127-2 slow-blow fuses (P/N 954773).
Thermo Fisher Scientific recommends
always replacing both fuses.
7. Reinstall the fuse cartridge.
8. Reconnect the main power cord to its
source and to the DP/SP.
9. Turn on the main power switch.
10. Press the
11. Reconnect the pump to the software by clicking Connected on the Pump
panel.
12. Turn on the pump flow.
Doc. 065342-06 12/11
POWER
button on the front of the DP/SP for 2 seconds to
2-7).
Figure
POWER
button on the front of the DP/SP to turn on the pump.
9-13).
Figure 9-13. DP/SP Fuse Cartridge