4. Using tweezers, remove the PdH reference electrode gasket from the well
(see
Figure
Figure 9-68. Removing the PdH Reference Electrode Gasket
5. Using tweezers, grasp the new PdH reference electrode gasket
(P/N 072214) on its edge (see
gasket cutout, do not place the tweezer tips on the cutout.
Figure 9-69. Installing the PdH Reference Electrode Gasket
6. Align the gasket so that the notched edge of the gasket aligns with the
corresponding notch in the reference electrode well and press the gasket
into the well.
402
9-68).
Notch in
well
Cutout
PdH Reference
Electrode Gasket
Figure
9-69). To avoid deforming the
PdH Reference
Electrode Gasket
Doc. 065342-06 12/11