Thermo Scientific Dionex ICS-5000 Operator's Manual page 106

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 Operator's Manual
ED detectors 1 and 2 (on/off)
CD/ED detectors 1 and 2 auto-offset
CD/ED detectors 1 and 2 mark
Suppressors 1 and 2 (on/off)
Reaction coil heater (on/off)
A/C relays 1 and 2 (open/closed)
Assigning TTL Input Functions
TTL input control functions are assigned in the Chromeleon 7 Instrument
Configuration program or the Chromeleon 6.8 Server Configuration program.
You can assign one or more functions to each input.
NOTE
If two timebases are sharing the DC, the TTL input functions
are shared between the timebases. For example, if TTL input
1 is assigned to control injection valve 1, that function is
assigned to both timebases. You do not need to associate a
TTL input with a particular timebase.
To select TTL input control functions:
1. Open the Chromeleon 7 Instrument Configuration program or
Chromeleon 6.8 Server Configuration program.
2. Double-click the DC icon under the instrument or timebase.
90
Doc. 065342-06 12/11