Dionex ICS-5000 DC Detector/Chromatography Compartment Specifications
Automation Support of
Third-Party Instruments
Customizable System
Control Panels
Signal Channels
Data Trending Plots
System Status Virtual
Channels
Power Failure Protection
System Trigger
Commands and
Conditionals
Daily Audit Trail
Sample Audit Trail
System Calibration
Storage
Customized Reporting
GLP Compliance
Electrical
Main Power
Requirements
Fuse Requirements
426
Yes, fully controls over 300 different instruments
from more than 30 manufacturers, including GC,
HPLC, and MS.
Yes, standard feature
Detector signals, detector background signals,
temperatures
Yes, all device numerical parameters plotted
Yes, standard feature
Yes, standard feature
Yes, standard feature
Yes, standard feature
Yes, standard feature
Yes, factory, current, and previous. Completely
user selectable
Yes, standard feature with unlimited report
workbooks
Yes, optional
90–265 VAC, 47–63 Hz (Auto-sensing power
supply; no manual voltage or frequency adjustment
required)
Typical input power: 200 W
Maximum line draw: 9.2 A at 110 VAC
Two IEC 60127-2 slow-blow fuses (P/N 954772)
rated at 10 A, 250 V