NOTE
For routine on/off control, use the
of the DP/SP (see
pump, press and hold the
The fuse cartridge contains two 2-amp IEC 60127-2 slow-blow fuses
(P/N 954773). For instructions on how to change the fuses, see
The power cord plugs into the IEC 320 three-prong receptacle.
The power supply cord is used as the main disconnect device. Make
sure the socket-outlet is located near the DC and is easily accessible.
Le cordon d'alimentation principal est utilisé comme dispositif principal
de débranchement. Veillez à ce que la prise de base soit située/installée
près du module et facilement accessible.
Das Netzkabel ist das wichtigste Mittel zur Stromunterbrechung. Stellen
Sie sicher, daß sich die Steckdose nahe am Gerät befindet und leicht
zugänglich ist.
Digital I/O Port
The digital I/O port provides a connection to auxiliary devices. The connector
includes three TTL inputs and four relay outputs.
functions assigned to the connector pins.
The maximum switching voltage of the relays is 24 V. The switching
current must not exceed 100 mA.
Pin Number
1
2
3
4
5
6
7
Doc. 065342-06 12/11
Figure 2-1
and
POWER
Signal Name
Signal Level
-----
-----
-----
-----
Relay 3 Out
Potential-free
Relay 1 Out
Potential-free
Relay 2 Out
Potential-free
Relay 3 Out
Potential-free
Relay 1 Out
Potential-free
Table 2-1. 25-Pin D-Sub I/O Port (Female)
2 • Description
POWER
button on the front
Figure
2-2). To turn off the
button for 2 seconds.
Table 2-1
indicates the
Description
Not used
Not used
Normally open
Normally closed
Normally closed
Normally closed
Common
Section
9.11.
31