Thermo Scientific Dionex ICS-5000 Operator's Manual page 412

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 Operator's Manual
Installing the pH-Ag/AgCl Reference Electrode in the Cell for a Capillary
Capillary
IC
IC System
1. To avoid any hydraulic pressure buildup when inserting the pH-Ag/AgCl
reference electrode into the cell, make sure that fitting plugs are not
installed on the cell inlet and outlet fittings.
2. Use a sharp tool (for example, the tip of a safety pin or a straight pin) to
remove the pH-Ag/AgCl reference electrode O-ring (see
3. Using tweezers, install the reference electrode gasket for capillary IC
(P/N 072162) into the bottom of the reference electrode well (see
Figure
396
Figure 9-59. pH-Ag/AgCl Reference Electrode O-Ring
9-60).
Figure 9-60. Reference Electrode Gasket for Capillary IC
Figure
9-59).
pH-Ag/AgCl
Reference
Electrode
O-Ring
(remove for
capillary IC)
Reference
Electrode Gasket
for Capillary IC
Doc. 065342-06 12/11