Thermo Scientific Dionex ICS-5000 Operator's Manual page 292

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 Operator's Manual
All samples scheduled for run on timebases x need to specify the
same value for property y.
When a TC is shared by two timebases, the same values must be selected for
certain parameters. If the Chromeleon Ready Check detects a discrepancy,
this error occurs and batch processing cannot be started.
To troubleshoot:
Find and eliminate any discrepancies in the parameter settings. For example,
all samples must turn on TempCtrl or all samples must turn off TempCtrl.
Can't execute "Disconnect" command during data acquisition.
-or-
Can't execute "Standby" command during data acquisition.
Selecting the Disconnect or Standby command during data acquisition will
stop data transmission and cause data acquisition to stall. If you attempt to
issue one of these commands during data acquisition, this error occurs.
To troubleshoot:
Wait until data acquisition is completed before attempting to issue either of
these commands.
Can't find download image TC3000.HEX.
TC3000.HEX is the version of the TC instrument control firmware file that
should be available in the \Bin directory of your Chromeleon installation. This
error occurs if the file is missing.
To troubleshoot:
Select Tools > Instrument Qualification (Chromeleon 7) or Qualification >
Chromeleon 6.8 IQ (Chromeleon 6.8) to run the Chromeleon IQ and
diagnose the problem. When you eliminate the source of the error, rerun the
Setup program.
276
Doc. 065342-06 12/11