7. To avoid any hydraulic pressure buildup when inserting the reference
electrode into the cell, make sure that fitting plugs are not installed on the
cell inlet and outlet fittings.
8. Align the fitting on the end of the PdH reference electrode so that the
knobs on the fitting align with the grooves in the reference electrode well
(see
Figure
9. Insert the fitting into the well.
Figure 9-70. Inserting the PdH Reference Electrode Fitting into the
10. Screw the nut on the PdH reference electrode into the reference electrode
well and tighten it fingertight (see
use a wrench to tighten the nut an additional 20 to 30 degrees.
Doc. 065342-06 12/11
9-70).
PdH Reference
Electrode
Grooves
Reference Electrode Well
Nut
Figure 9-71. Installing the PdH Reference Electrode Nut
Knobs
Figure
9-71). After fingertightening,
9 • DC Service