Thermo Scientific Dionex ICS-5000 Operator's Manual page 433

Ion chromatography system
Table of Contents

Advertisement

ICS-5000 SP Single Pump and DP Dual Pump Specifications (Continued)
System Trigger
Commands and
Conditionals
Daily Audit Trail
Sample Audit Trail
System Calibration
Storage
Customized Reporting
GLP Compliance
Electrical
Main Power
Requirements
Fuse Requirements
Physical
Dimensions (h × w × d)
Weight
Flow Path
Environmental
Operating Temperature
Operating Humidity
Doc. 065342-06 12/11
Yes, standard feature
Yes, standard feature
Yes, standard feature
Yes, factory, present, and previous. Completely
user selectable
Yes, standard feature with unlimited report
workbooks
Yes, optional
90–264 VAC, 47–63 Hz
(Auto-sensing power supply; no manual voltage or
frequency adjustment required)
Typical input power: 28 W
Maximum line draw: 1.2 A at 110 VAC
Two IEC 60127-2 slow-blow fuses (P/N 954773)
rated at 2 A, 250 V
36 × 21 × 48 cm (14 × 8.25 × 19 in.)
SP: 20.4 kg (45 lb)
DP: 24.1 kg (55 lb)
All polymeric (PEEK), anion or cation
configurations
4–40 °C (40–104 °F)
5–95% relative humidity (noncondensing)
A • Specifications