d. Pull on the fitting slightly to remove the PharMed tubing from the
upper notch on the pump mounting plate.
5. Pull off the PharMed and Tygon 2075 tubing from the fitting. Save the
fitting.
6. Pull off all seal wash tubing from the seal wash reservoir and the primary
and secondary pump heads(see
7. Cut three new pieces of Tygon 2075 tubing (0.15 cm (0.060 in) ID;
P/N 064079) and one piece of PharMed tubing (0.159 cm (0.0625 in) ID;
P/N 063268) to the same lengths as the tubing just removed. The tubing is
included in the pump ship kit:
Pump Ship Kit
DP Ship Kit Dual Capillary
DP Ship Kit Dual Analytical
DP Ship Kit Capillary and Analytical
SP Ship Kit Capillary or Analytical
Doc. 065342-06 12/11
1
6
2
5
3
4
Figure 9-9. Peristaltic Pump for Seal Wash System
Figure
9 • DP/SP Service
PharMed Tubing
1
(P/N 063268)
Rotor
2
Lever
3
Mounting Plate
4
Tygon 2075 Tubing
5
(P/N 064079)
Fitting
6
9-10).
Part Number
072112
062463
072111
063342
327