Thermo Scientific Dionex ICS-5000 Operator's Manual page 202

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 Operator's Manual
2. Verify that the current to the EGC and the suppressor is off.
Supplying power to the EGC when there is no eluent flow can severely
damage the cartridge.
3. If you plan to store the EGC in the EG, no further action is required.
If you plan to store the cartridge outside the EG (for example, in the
original shipping container), follow the instructions in
remove the EGC from the EG, and the instructions in
store the EGC.
To restart the EG:
1. Select the parameters for the analysis on the Chromeleon 7 ePanel Set or
the Chromeleon 6.8 panel tabset. For EG startup information, see
Section
2. Condition the EGC. For instructions, see
EGC" on page 343
3. Hydrate the CR-TC (see
page
350).
4. Hydrate the suppressor as instructed in the product manual. Suppressor
manuals are included on the Thermo Scientific Reference Library DVD
(P/N 053891).
5. Let the system equilibrate before collecting data.
Before shipping the EG:
1. Follow the instructions in
EG.
2. Plug all fittings. Tighten the vent fitting (capillary EGC) or cap the vent
port (analytical EGC). This will prevent the evaporation of water in the
electrolyte reservoir.
Make sure the EGC packaging conforms to hazardous material
shipping requirements. Refer to the Material Safety Data Sheet (MSDS)
shipped with the EGC for the chemical description.
186
4.9.
or
"Conditioning the Analytical EGC" on page
"Installing and Flushing the New CR-TC" on
Section
Section
Section
"Conditioning the Capillary
9.14.1, to remove the EGC from the
9.14.1, to
9.14.3, to
348.
Doc. 065342-06 12/11