The injections limit for the column has been exceeded.
-or-
The injections limit will be exceeded for the column x.
Chromeleon monitors the number of injections done with each column and
display an appropriate message (shown above) at the following times: when
the injections limit is about to be exceeded or has been exceeded.
To troubleshoot:
Replace the column.
The nominal temperature must be set within the current limits.
This error occurs if the selected TC temperature setting is outside the range
specified by the Temperature.LowerLimit and Temperature.UpperLimit
commands.
To troubleshoot:
Select a valid temperature setting.
The nominal temperature value is out of range defined by the
active column ID.
This error occurs if the selected TC temperature setting is outside the range
specified by the Column_x.Temp_LowerLimit and Column_x.Temp_
UpperLimit commands.
To troubleshoot:
Select a valid temperature setting.
The oven is not fully operable yet. Check the values of "Standby,"
"Ready," "Door," and "CardState."
Each time the TC power is turned on, the firmware runs a series of self-
diagnostics in which the status of the main components is checked. This error
occurs if a component is not ready for operation.