d. Connect the
the waste line to the CR-TC
(0.0025 in) ID
PEEK Tubing
(P/N 072204)
3. If you are installing a CR-TC in an analytical IC system:
a. Locate the CR-TC Hydration Tubing Kit (P/N 063487) provided in
the EG Ship Kit.
Doc. 065342-06 12/11
EGC OUT
line to the CR-TC
EGC
(Capillary)
0.062 mm
Assembly
EGC
OUT
To CR-TC
Regen Out
(blue label)
To CR-TC
Eluent In
(red label)
To CR-TC
Regen In
(orange label)
Figure 9-26. Plumbing for Hydrating a Capillary CR-TC
ELUENT IN
REGEN OUT
port.
Capillary CR-TC Flushing:
Pump at 0.02 mL/min for 15 min
EGC
IN
ELUENT
IN
TO PUMP/DAMPER
1.58 mm (0.062 in) ID
To CR-TC
Tefzel Tubing
Eluent Out
Assembly (P/N 072225)
(yellow label)
9 • EG Service
port and connect
0.25 mm (0.010 in)
ID
PEEK Tubing
(Temporary
connection)
351