Thermo Scientific Dionex ICS-5000 Operator's Manual page 296

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 Operator's Manual
Communication time-out.
This error occurs if a TC firmware command is not executed within the time
allotted for its completion.
To troubleshoot:
1. Check the USB connections.
2. Check the connection from the TC to the power supply.
3. If you cannot identify the cause of the error, contact Technical Support for
Dionex products for assistance.
Configuration doesn't match. X not installed.
-or-
Configuration mismatch (x - CM uses y, Column Compartment
uses z). Please use the Server Configuration program to check
the installation.
This error occurs if there is a discrepancy between the TC properties and the
actual configuration.
To troubleshoot:
1. Start the Chromeleon 7 Instrument Configuration Manager or the
Chromeleon 6.8 Server Configuration program.
2. In the instrument or timebase, right-click the TC and select Properties.
3. Correct any discrepancies in the Properties dialog box.
Current flow is out of limits defined for the column.
This error occurs if the current flow rate exceeds the value specified by the
Column_x.FlowRate_UpperLimit command.
To troubleshoot:
Select a valid flow rate.
280
Doc. 065342-06 12/11