Thermo Scientific Dionex ICS-5000 Operator's Manual page 57

Ion chromatography system
Table of Contents

Advertisement

Eluent Generator Cartridge (EGC)
Table 2-2
lists the versions of eluent generator cartridges available for use
with the EG. Each EGC contains an electrolyte concentrate solution
appropriate for the particular eluent generation. Each EGC is designed to be
used in either a capillary IC system or an analytical IC system. An EGC
designed for a capillary IC system cannot be used in an analytical IC system,
and vice versa.
Analytical IC systems require EGC III cartridges and (if used) the EPM III
Electrolytic pH Modifier. Previous versions of EGC or EPM are not
compatible with the Dionex ICS-5000 Eluent Generator.
EGC Version
Eluent Generator Cartridges for Capillary IC Systems
EGC KOH (Capillary)
EGC MSA (Capillary)
Eluent Generator Cartridges and Accessories for Analytical IC Systems
EGC III K
2
EPM III Electrolytic
pH Modifier
EGC III KOH
EGC III LiOH
EGC III MSA
EGC III NaOH
Doc. 065342-06 12/11
Part
Number
072076
072077
CO
074536
3
080135
074532
074534
074535
074533
Table 2-2. Capillary and Analytical EGCs
2 • EG Description
Function
Generates potassium hydroxide eluent
for anion exchange separations.
Generates methanesulfonic acid eluent
for cation exchange separations.
Generates potassium carbonate eluent
for anion exchange separations.
Produces a carbonate/ bicarbonate
mixture when installed with the EGC III
K
CO
. Requires a carbonate mixer:
2
3
P/N 061686 (4 mm); P/N 063443
(2 mm).
Generates potassium hydroxide eluent
for anion exchange separations.
Generates lithium hydroxide eluent for
anion exchange separations.
Generates methanesulfonic acid eluent
for cation exchange separations.
Generates sodium hydroxide eluent for
anion exchange separations.
41