Thermo Scientific Dionex ICS-5000 Operator's Manual page 425

Ion chromatography system
Table of Contents

Advertisement

4. Select the check box for the column to be monitored (A, B, C, or D) and
click OK.
To install a column ID chip card:
1. To attach a column ID chip card to the column,
wrap the strap around the column, insert the
rivet into one of the holes on the strap, and
press to secure the strap to the column (see
Figure
2. Insert the chip card (with the Dionex logo
facing up) into the appropriate card reader (A,
B, C, or D) along the top of the compartment
(see
Figure
When the chip card is installed correctly, the
LED next to the slot is green.
3. Install a chip card for any other columns to be monitored.
To select the column properties to be monitored:
1. Close the TC door, if it is open. (You cannot select column properties
unless the door is closed.)
2. On the TC Control panel, click the Column Details... button.
This opens the Column Details screen.
3. For each column, specify the properties the column ID system will
monitor. When you finish, click Close.
Doc. 065342-06 12/11
9-73).
Figure 9-73. Attaching the Column ID System to the Column
9-74).
Figure 9-74. Inserting a Column ID Chip Card
9 • TC Service
409