EGC is either not installed or invalid cartridge number. Please
install an EGC.
This error occurs if no EGC cartridge is connected to the EG or the cartridge
serial number is not valid for the type of IC system.
To troubleshoot:
1. Verify that an EGC of the correct type (capillary or analytical) is installed.
2. Verify that the EGC cable is securely plugged into the EG (see
Figure
Leak sensor wet.
This error occurs when liquid accumulates in the drip tray in the bottom of the
EG.
To troubleshoot:
1. Locate the source of the leak by visually inspecting the tubing, fittings,
and components in the EG.
2. Tighten fittings (or replace tubing and fittings) as required (see
Section
various types of leaks.
3. After fixing the leak, dry the drip tray and leak sensor thoroughly to
prevent the leak sensor from triggering additional error messages.
pH modifier must be used with a different type of cartridge.
This error occurs if you attempt to configure an EPM III Electrolytic pH
Modifier with an incompatible type of EGC.
To troubleshoot:
Verify that the EPM III is paired with an EGC III K
be used with any other cartridge type.
252
2-10) and the locking ring is tightened.
9.12). Refer to
Section 8.22
for detailed troubleshooting of
CO
. The EPM III cannot
2
3
Doc. 065342-06 12/11