Tubing Size and Type
10 mm (0.39 in) ID
polyethylene
1/16 in fitting bolt
1/16 in ferrule
1/8 in fitting bolt
1/8 in ferrule
High-pressure 10-32
fitting bolt
High-pressure 10-32
ferrule, double-cone
10-32 fitting bolt
10-32 ferrule, double-
cone
Notes
For tightening requirements for the 10-32 fitting bolt (P/N 043275) and
ferrule (P/N 043276), refer to Installation of Dionex Liquid Line Fittings
(Document No. 031432). The manual is on the Thermo Scientific
Reference Library DVD (P/N 053891).
For tightening requirements for the high-pressure fitting bolt
(P/N 074449) and ferrule (P/N 074373), see
Use a tubing cutter to cut tubing to the required length. Make sure the cut
is at a right angle to the length of the tubing and there are no nicks or burrs
on the end. A tubing cutter (P/N 049584) is included in the DC Ship Kit
(P/N
072011). Refer to the instructions provided with the cutter.
Doc. 065342-06 12/11
Color
Colorless
Tan
Tan
Tan
Yellow
Blue
Blue
Black
Tan
Table 9-1. DP/SPTubing and Fittings (Continued)
9 • DP/SP Service
P/N
Used to connect...
055075
Drip tray drain hose
052230
Inlet check valve fitting
062511
052267
Eluent reservoir fittings
048949
074449
Capillary pumps only: Pulse
damper to EGC or injection
valve
074373
Use with high-pressure fitting
bolts
043275
All other tubing fittings
043276
Section