Thermo Scientific Dionex ICS-5000 Operator's Manual page 285

Ion chromatography system
Table of Contents

Advertisement

2. Rehydrate the suppressor. Refer to the suppressor manual for instructions.
Suppressor manuals are included on the Thermo Scientific Reference
Library DVD (P/N 053891).
3. If the error persists, replace the suppressor (see
Suppressor over-voltage.
This error appears if you turn on the suppressor and the system cannot
establish a connection with the suppressor.
To troubleshoot:
1. Check the suppressor cable connection.
2. If the error persists, replace the suppressor (see
Suppressor stopped for flow rate.
This message occurs if the pump flow stops while the suppressor is on. The
suppressor is automatically turned off to prevent damage to the suppressor.
To troubleshoot:
If the pump stopped unexpectedly, see
information.
Upper door opened.
This message occurs when the door to the upper compartment is opened
during a run.
To troubleshoot:
1. Verify that the door is fully closed.
2. Check for—and remove—any obstruction.
3. If the door is fully closed and the error persists, contact Technical Support
for Dionex products for assistance.
Doc. 065342-06 12/11
8 • DC Troubleshooting
Section
Section
Section 8.14
for pump troubleshooting
9.27.2).
9.27.2).
269