3. Check that the main power cord is plugged into both the EG rear
panel connector and the power source. Check that the wall outlet has
power.
•
Injection valve is not actuating or the sample loop is plugged
1. Test the valve by manually switching the position from the DC ePanel
in Chromeleon 7 or Control panel in Chromeleon 6.8.
2. Check the sample loop for blockage. Clean or replace the loop if
needed.
•
Detector not properly installed
CD: An electronics connector on the back of the detector plugs into a
receptacle on the detector compartment. Push on the detector to make
certain the connector is securely connected (see
•
ED: Verify that the signal cables from the cell are connected to the
detector block. In addition, an electronics connector on the back of the
detector plugs into a receptacle on the detector compartment. Push on the
detector to make certain the connector is securely connected (see
Figure
ED Only
•
Cell is off
Turn on the cell from the detector ePanel (in Chromeleon 7) or Control
panel (in Chromeleon 6.8).
Doc. 065342-06 12/11
8-1).
Push here to ensure that the
detector's electrical
connection is secure.
Figure 8-1. Detector Connection
8 • Troubleshooting
Figure
8-1).