3. Check that the main power cord is plugged into both the EG rear
panel connector and the power source. Check that the wall outlet has
power.
Injection valve is not actuating or the sample loop is plugged
1. Test the valve by manually switching the position from the DC ePanel
in Chromeleon 7 or Control panel in Chromeleon 6.8.
2. Check the sample loop for blockage. Clean or replace the loop if
needed.
Detector not properly installed
CD: An electronics connector on the back of the detector plugs into a
receptacle on the detector compartment. Push on the detector to make
certain the connector is securely connected (see
ED: Verify that the signal cables from the cell are connected to the
detector block. In addition, an electronics connector on the back of the
detector plugs into a receptacle on the detector compartment. Push on the
detector to make certain the connector is securely connected (see
Figure
ED Only
Cell is off
Turn on the cell from the detector ePanel (in Chromeleon 7) or Control
panel (in Chromeleon 6.8).
Doc. 065342-06 12/11
8-1).
Push here to ensure that the
detector's electrical
connection is secure.
Figure 8-1. Detector Connection
8 • Troubleshooting
Figure
8-1).