6. Press the release lever located below the detector. The detector pops out
slightly from the DC compartment. Grasp the detector and pull outward to
remove the detector from the compartment.
7. Note the electrical plug on the back of the detector and the receptacle on
the inside of the DC.
8. Push the new detector into the opening in the DC and press firmly to
ensure the electrical connection is secure.
9. Reinstall the cell and reconnect the cell inlet and outlet lines and the
cables.
NOTE
If you are installing an ED detector in a system that previ-
ously was set up for conductivity detection, remove the sup-
pressor cartridge from the IC Cube and replace it with a
suppressor bypass cartridge (P/N 072055). For IC Cube car-
tridge installation instructions, see
Doc. 065342-06 12/11
9 • DC Service
Section
9.26.1.