Preventive Maintenance Procedure for
Dionex ICS-1000, ICS-1100, ICS-1500,
ICS-1600, ICS-2000, and ICS-2100
Ion Chromatography Systems
Document No. 031928
Revision 05
October 2012

Advertisement

loading

Summary of Contents for Thermo Scientific Dionex ICS-1000

  • Page 1 Preventive Maintenance Procedure for Dionex ICS-1000, ICS-1100, ICS-1500, ICS-1600, ICS-2000, and ICS-2100 Ion Chromatography Systems Document No. 031928 Revision 05 October 2012...
  • Page 2 2012 Thermo Fisher Scientific Inc. All rights reserved. © Chromeleon is a registered trademark of Thermo Fisher Scientific Inc. in the United States. Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
  • Page 3: Preventive Maintenance Instructions

    Preventive Maintenance Instructions These instructions describe a preventive maintenance procedure for the Thermo ™ Fisher Scientific Dionex ICS-1000, ICS-1100, ICS-1500, ICS-1600, ICS-2000, and ICS-2100 Ion Chromatography Systems. Thermo Fisher Scientific recommends performing this procedure annually, as well as before scheduled Performance Qualification tests.
  • Page 4 Preventive Maintenance Instructions Substitution of non-Thermo Scientific/Dionex parts may impair system performance and void the product warranty. For details, see the warranty statement in the Dionex Terms and Conditions. Additional Items Required The following items are provided in the Ship Kit for each system: •...
  • Page 5: Rebuilding The Injection Valve

    Dionex ICS-1000, ICS-1100, ICS-1500, ICS-1600, ICS-2000, and ICS-2100 DO NOT OVERTIGHTEN FITTINGS. Overtightening a fitting can restrict the tubing ID, which can cause damage to the system. The split-cone ferrules shown here are especially sensitive to overtightening. Tighten these fittings fingertight only and then tighten them slightly more only if they leak.
  • Page 6 Preventive Maintenance Instructions Figure 2. Dionex ICS-2000 and ICS-2100 Injection Valve Plumbing NOTE A backpressure coil may be installed between port P and the line labeled TO INJECT VALVE - P 6. Close the door. 7. Turn on the pump. Record the completed steps on the Preventive Maintenance Checklist.
  • Page 7 3). To open the valve, turn the knob one-quarter to one-half turn counterclockwise. 2. After rinsing, close the waste valve and turn off the pump. 3. Close the eluent valve from the system Control panel in the Thermo Scientific ™ Dionex Chromeleon...
  • Page 8 Preventive Maintenance Instructions NOTE The inlet check valve assembly housing has a 1/4-28 port. The outlet check valve assembly housing has a 10- 32 port (see Figure 4. Place a new cartridge (P/N 045994) in Inlet Check Valve Outlet Check Valve the inlet check valve housing with the 1/4-28 Port 10-32 Port...
  • Page 9 Dionex ICS-1000, ICS-1100, ICS-1500, ICS-1600, ICS-2000, and ICS-2100 Replacing Pump Piston Rinse Seals and Piston Seals Note About Replacing a Pump Piston Pump piston replacement is not part of the preventive maintenance procedure. However, if replacement is required (for example, because the piston is dirty,...
  • Page 10 Preventive Maintenance Instructions Priming Valve Knob (P/N 055709) Outlet Check Valve Assembly (P/N 045721) O-Ring (P/N 055752) Piston Seal (P/N 055870) Spacer with Drain (P/N 057913) Primary Pump Head (P/N 055711) Piston Rinse Seal (P/N 048722) O-Ring (P/N 014895) O-Ring Inlet Check (P/N 059283) Valve Assembly...
  • Page 11 Dionex ICS-1000, ICS-1100, ICS-1500, ICS-1600, ICS-2000, and ICS-2100 6. The piston does not come off as part of the pump head assembly because it is captured by a magnetic retention system. To remove the piston after removing the pump head: hold the shaft of the piston (near the base), tilt the piston slightly, and pull the piston away from the pump.
  • Page 12 Preventive Maintenance Instructions Removing the Piston Seal from the Head 1. If this is the primary pump head, install a 10-32 fitting plug (P/N 042772) on the outlet check valve. Tighten the plug. 2. If this is the secondary pump head, install a 10-32 fitting plug (P/N 042772) in both the inlet and outlet ports.
  • Page 13 Dionex ICS-1000, ICS-1100, ICS-1500, ICS-1600, ICS-2000, and ICS-2100 Reinstalling the Head and Piston Thermo Fisher Scientific recommends reinstalling the head and piston as a single assembly, so that the piston centers itself onto the magnetic follower. 1. Hold the assembled spacer and guide with the drain tubes aligned vertically and press the spacer into the head until it is flush with the indented surface of the head.
  • Page 14 Preventive Maintenance Instructions 2. If the O-ring is removed with the valve knob in Step 1, pull the O-ring off the end of the knob (see Figure 9). If the O-ring is not removed with the valve knob, insert a thin object (for example, the bent end of a paper clip) into the cavity in the pump head and carefully extract the O-ring.
  • Page 15 Dionex ICS-1000, ICS-1100, ICS-1500, ICS-1600, ICS-2000, and ICS-2100 Replacing the End-Line Filter Install the new end-line filter (P/N 045987) on the end of the eluent or deionized water line, inside the reservoir. Verify that the end of the filter extends to the bottom of the reservoir and that the filter is submerged in liquid.
  • Page 16: Preventive Maintenance Checklist

    Preventive Maintenance Instructions Preventive Maintenance Checklist Dionex Product Serviced:       ICS-1000 ICS-1100 ICS-1500 ICS-1600 ICS-2000 ICS-2100 1. Injection valve rebuild (see Section 1) 5. Waste valve and priming valve O-ring replacement (see Section   Rebuilt injection valve Installed new O-ring in waste valve ...