4. Click the Calibration button. The Wellness panel opens, see
Figure 9-58. Wellness Panel: Dionex ICS-5000 Electrochemical
5. Follow the instructions on the panel to calibrate the electrode.
6. Disconnect the pH-Ag/AgCl reference electrode cable and the cell cable.
7. Grasp the cell body and pull straight out to remove the cell from the
detector.
If you are installing the pH-Ag/AgCl reference electrode in a capillary IC
Capillary
IC
system, go on to
If you are installing the pH-Ag/AgCl reference electrode in an analytical IC
system, go on to
Doc. 065342-06 12/11
Detector
(Chromeleon 7 Version Shown)
page
396.
page
398.
9 • DC Service
Figure
9-58.
395