Thermo Scientific Dionex ICS-5000 Operator's Manual page 172

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 Operator's Manual
Setting Up the Seal Wash System (All pumps)
1. Add ASTM Type I (18 megohm-cm) filtered and deionized water that
meets the specifications listed in
(P/N 064155).
The liquid level should be between the Min. and Max. markers on the
reservoir label. Do not overfill the reservoir.
2. Place the cap on the reservoir and tighten fingertight.
3. Insert the reservoir into the holder on the component panel.
4. Check that the tubing is engaged in the peristaltic pump (see
If it is not, lift the lever on the peristaltic pump up and to the right, insert
the tubing neatly between the lever and the rotor, and release the lever to
secure the tubing.
Pump Head
Outlet Tubing
from Seal Wash
5. Follow these steps to activate the piston seal wash function:
a. Open the Chromeleon 7 ePanel Set or Chromeleon 6.8 panel tabset.
b. Press the F8 key to open the Commands dialog box.
c. Select the pump name.
d. For Chromeleon 7, click the Properties tab on the right pane of the
dialog box. For Chromeleon 6.8, scroll through the list of commands
and properties under the pump name.
156
Tubing to
Secondary
Reservoir
Figure 4-6. Peristaltic Pump
Section 1.5
to the seal wash reservoir
Rotor
Figure
4-6).
Lever
Doc. 065342-06 12/11