Dionex ICS-5000 Operator's Manual
6. Remove the guard column (if present)
and separator column from the column
clips. Lift up the separator column
outlet fitting (see
to free it from the clips and then slide it
out of the cartridge.
7. Remove the columns from the cartridge.
8. Disconnect the tubing from the
columns. Save the tubing.
NOTE
9. Before installing the new separator column, tear off the column label and
slide it into the holder on the front of the cartridge (see
10. If the new column configuration is the same as the previous, connect the
precision cut tubing that you removed in
376
Figure
9-44) slightly
Figure 9-44. Separator Column Outlet Fitting
Do not cut the tubing that you removed from the col-
umns and do not substitute different tubing. To ensure
good chromatographic results, capillary tubing is cut
at the factory using a specialized precision cutter. If
you need to replace the tubing or to install a different
length tubing (to accommodate a new column configu-
ration), order an IC Cube Tubing Kit (P/N 072186).
Figure
Step 8
to the new columns.
Doc. 065342-06 12/11
9-48).