2 • DC Description
contains the plumbing connections necessary for eluent and waste flow
through the system.
Suppressor Cartridge or Suppressor Bypass Cartridge
Capillary
IC
Two versions of capillary suppressor cartridges are available: ACES 300
Anion Capillary Electrolytic Suppressor cartridge (P/N 072052) and CCES
300 Cation Capillary Electrolytic Suppressor cartridge (P/N 072053). Each
cartridge holds a suppressor and the required tubing connections for eluent
and regenerant flow to the suppressor. The suppressor and tubing connections
inside the cartridge are not accessible. Tubing exits the front of the cartridge
to allow connection to other capillary IC components.
When the cartridge is installed in the IC Cube, pins on the rear of the cartridge
connect the suppressor to its power source. For details about the suppressors,
refer to the suppressor manuals. The manuals are on the Thermo Scientific
Reference Library DVD (P/N 053891).
Installation of a suppressor is optional. However, if one is not used, a
suppressor bypass cartridge (P/N 072055) must be installed. The suppressor
bypass cartridge contains the plumbing connections necessary for eluent and
waste flow through the system.
57
Doc. 065342-06 12/11