Notes for Generating Carbonate/Bicarbonate Eluent
If an EGC III K
(EGC_1 and EGC_2, respectively):
1. Set EGC_1 Target Concentration to the concentration of K
required for your application
2. Set EGC_2 Target Concentration to the concentration of KHCO
required.
The K
CO
2
3
the EPM modifies the total to achieve the K
you require.
For example:
For a 3.50 mM K
EGC_2 to 1.00 mM.
The K
CO
2
3
Concentration field for the K
K
CO
eluent to achieve the required 3.50 mM K
2
3
mixture.
164
CO
and an EPM III Electrolytic pH Modifier are installed
2
3
cartridge generates the total of the two target concentrations and
CO
/1.00 mM KHCO
2
3
cartridge generates 4.50 mM K
CO
2
CO
/KHCO
2
3
eluent, set EGC_1 to 3.50 mM and
3
CO
(indicated in the Applied
2
3
EGC). The EPM modifies the 4.50 mM
3
CO
/1.00 mM KHCO
2
3
CO
2
3
3
eluent mixture that
3
3
Doc. 065342-06 12/11