Thermo Scientific Dionex ICS-5000 Operator's Manual page 242

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 Operator's Manual
(DC Amperometry and Integrated Amperometry modes)
Amperometric detection cell working electrode shorted to counter
electrode
Clean the working electrode with water and dry with pressurized air or
nitrogen.
Remove any precipitate on the counter electrode by cleaning the spot
directly opposite the working electrode with a lint-free tissue.
(DC Amperometry and Integrated Amperometry modes)—Leak
between gasket and electrode, or between gasket and cell body
Remove any precipitate with water and a lint-free tissue and then install a
new gasket (see
Turn the yoke knob (see
engage the knob.
226
Section
9.28.3).
Figure
9-53) a full 360 degrees to completely
Doc. 065342-06 12/11