To monitor the pH-Ag/AgCl reference electrode pH:
1. At installation, calibrate the pH electrode (see
2. When you run your first chromatographic instrument method or program,
note the pH value displayed on the Chromeleon panel (see
Figure 4-1. Monitoring the pH-Ag/AgCl Reference Electrode pH
3. Thereafter, monitor the pH value to determine whether there is a shift in
the pH. A shift in the pH reading for the same eluent composition
indicates a change in the Ag/AgCl reference potential.
If the pH value shifts by 0.5 pH unit from the value first observed, check
the pH-Ag/AgCl reference electrode (see
144
Section
9.28.6).