Dionex ICS-5000 Operator's Manual
11. Remove the fitting plug from the outlet fitting on the separator column.
Orient the fitting with a flat side up and push the fitting into the opening at
the front of the column cartridge until it stops (see
12. Coil the separator column tubing inside the cartridge as shown in
Figure 9-42
clips on the column cartridge.
13. Secure the inlet and outlet fittings on the guard column (if used) in the
column clips on the lid of the column cartridge.
14. Route the column inlet tubing from the guard column (if used) or the
separator column through the clip on the top edge of the cartridge lid.
15. Close the lid (you should hear a click) and route the tubing into the slot on
the front of the column cartridge (see
NOTE
3
16. Slide the column cartridge into the IC Cube column oven and tighten the
two thumbscrews.
17. Turn on the pump flow.
18. Wait until liquid is flowing out of the
valve.
378
or
Figure
9-43. Secure the column and the inlet fitting in the
If the columns are installed correctly, the cartridge lid
snaps closed easily. If the lid does not close easily, do
not force it. Open the lid and verify that the columns
and tubing are installed correctly and secured in the
clips.
2
1
Figure 9-48. Column Cartridge Closed
Figure
9-44).
Figure
9-48).
Separator Column
1
Outlet
Column Label
2
Column Inlet Tubing
3
Eluent Out
port (3) of the injection
Doc. 065342-06 12/11