Current pressure is out of limits defined for the column.
This error occurs if the current pressure reading is outside the range specified
by the Column_x.Pressure_UpperLimit and Column_x.Pressure_
LowerLimit commands.
To troubleshoot:
Find and eliminate any liquid leaks or blockages in tubing connections
(see
Section
Current temperature is out of limits defined for the column.
This error occurs if the current temperature reading is outside the range
specified by the Column_x.Temp_UpperLimit and Column_x.Temp_
LowerLimit commands.
To troubleshoot:
Select a valid temperature setting.
Data buffer overrun.
This error occurs in Chromeleon 6.8 if the system cannot write incoming data
to disk fast enough.
To troubleshoot:
Increase the buffer size:
1. Start the Chromeleon 7 Instrument Configuration Manager or the
Chromeleon 6.8 Server Configuration program.
2. Right-click the computer name and select Properties on the menu.
3. Click the Advanced tab and enter a new value (in seconds) in the Buffer
field. Click OK.
4. If the error occurs again, contact Technical Support for Dionex products
for assistance.
Doc. 065342-06 12/11
8.31). Adjust the flow rate accordingly.
8 • TC Troubleshooting
281