Thermo Scientific Dionex ICS-5000 Operator's Manual page 443

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 DC Detector/Chromatography Compartment Specifications
Physical Specifications
Dimensions
(h × w × d):
Weight
Flow Path
Environmental
Operating Temperature
Operating Humidity
Dionex ICS-5000 Conductivity Detector Specifications
Electronics Type
Cell Drive
Linearity
Resolution
Output Range
Noise, Wet
Filter
Sampling Rate
Cell Temperature
Cell Temperature
Stability
Cell Temperature
Compensation
Flow Cell Maximum
Pressure
Flow Cell Volume
Doc. 065342-06 12/11
44.5 × 42 × 57 cm (17.5 × 16.5 × 22.5 in)
38 kg (84 lb)
All polymeric (PEEK), anion or cation
configurations
4–40 °C (40–104 °F)
5– 95% relative humidity (noncondensing)
Microprocessor controlled digital signal processing,
autoranging
8 kHz square wave
1%
0.00238 nS/cm
Digital Signal Range: 0–15,000 µS/cm,
Analog Signal Range: 0–15,000 µS/cm
<0.2 nS at 23 µS/cm background
<0.1 nS at 1 µS/cm background
Rise times 0 to 10 s, programmable
1 to 100 Hz, user settable or automatic
5 °C above DC upper zone temperature to 60 °C
maximum. User settable, working range is identical
to settable range.
<0.001 °C
Default 1.7% per °C; programmable from 0–3% per
°C
10 MPa (1500 psi)
0.7 µL (Analytical), 0.02 µL (Capillary)
A • DC Specifications