Dionex ICS-5000 Operator's Manual
3
To troubleshoot:
1. If a sequence is being executed, terminate the sequence by selecting Stop
on the Chromeleon 7 ePanel or Chromeleon 6.8 Control panel.
2. Turn off the Thermo Scientific Dionex ICS-5000 Ion Chromatography
System power briefly by pressing the
module. Then press each button again to restart the system.
3. Try to toggle the valve from Load to Inject by pressing the
Inject
buttons on the front of the DC.
4. If the problem persists, contact Technical Support for Dionex products for
assistance.
IC Cube 1 heater over safe temperature.
-or-
IC Cube 2 heater over safe temperature.
This error occurs when the temperature of the IC Cube heater exceeds the
maximum allowed. This error may occur if the Thermo Scientific Dionex
ICS-5000 Ion Chromatography System is operating in an environment in
which the temperature is greater than 40 °C (104 °F).
To troubleshoot:
Refer to
Appendix A
266
4
Figure 8-7. High-Pressure Valves #3 and #4 (On AM)
for environmental specifications.
3
4
POWER
button on the front of each
Doc. 065342-06 12/11
High-
Pressure
Valve #3
High-
Pressure
Valve #4