Thermo Scientific Dionex ICS-5000 Operator's Manual page 171

Ion chromatography system
Table of Contents

Advertisement

1. Note the tubing connected from the peristaltic pump to the seal wash tube
on the secondary pump head on pump 1 (see
Disconnect this tubing from the pump head and connect it to the seal wash
tube on the secondary pump head on pump 2.
2. Disconnect the short piece of tubing between the primary and secondary
pump heads on pump 1 (see
the pump heads on pump 2.
3. Disconnect the seal wash reservoir inlet line from the primary pump head
on pump 1 (see
pump head on pump 2.
4. When you finish these connections, go to the next section to continue
setting up the seal wash system.
Doc. 065342-06 12/11
1
2
3
Figure 4-5. Piston Seal Wash Connections to Pump 1
Figure
4-5, item
Figure
4
2
Figure
4-5, item
). Connect this tubing to
3
). Connect the inlet line to the primary
4 • DP/SP Startup
1
4-5, item
).
Connection from
peristaltic pump to
1
secondary pump
head
Connection between
2
pump heads
Seal Wash Reservoir
3
Inlet Line
Seal Wash Reservoir
4
Outlet Line
155