Thermo Scientific Dionex ICS-5000 Operator's Manual page 308

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 Operator's Manual
Starting data acquisition manually will overwrite any data which
has been acquired so far.
This error occurs if you attempt to start data acquisition manually while a
batch is already running.
To troubleshoot:
To retain the data already acquired, cancel manual data acquisition; otherwise,
go ahead and start data acquisition.
The command can format only chip cards with an invalid data
structure.
This error occurs if you select the FormatChipCard command for a column
ID chip card that is not defective.
To troubleshoot:
Cancel the command. Do not attempt to reinitialize a chip card unless it
contains invalid or corrupted data.
The communication interface is closed. Reconnect the
instrument.
This error occurs if there is a communication failure between the TC and the
PC on which Chromeleon is installed.
To troubleshoot:
1. Check the USB connections.
2. Check the connection from the TC to the power supply.
3. If you cannot identify the cause of the error, contact Technical Support for
Dionex products for assistance.
292
Doc. 065342-06 12/11