Thermo Scientific Dionex ICS-5000 Operator's Manual page 294

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 Operator's Manual
b. Start the Chromeleon 7 Instrument Configuration Manager or the
Chromeleon 6.8 Server Configuration program.
c. In the instrument or timebase, right-click the TC for which you want
to update the firmware and select Properties on the context menu.
d. On the General tab page in the Properties dialog box, click
Firmware Download to display a file open dialog box.
e. Select the file (the extension is .fmw) containing the new firmware
version and click Open to begin the download. Audit Trail messages
will inform you of the status of the download.
3. If the error occurs again, contact Technical Support for Dionex products
for assistance. The firmware cannot be serviced by the user.
Can't use device "x" as a source of the column pressure. Please
check the property "SystemPressure" for a valid device name.
The device must have the "Pressure" property available.
-or-
Can't use device "x" as a source of the column pressure. Please
check the property "y" for a valid device name. The device must
have the "Pressure" property available.
This error occurs if the device selected to monitor the system pressure is
ineligible to do so.
To troubleshoot:
1. Open the Chromeleon 7 ePanel Set or the Chromeleon 6.8 panel tabset.
2. Press the F8 key to open the Commands dialog box.
3. Under TC, select the column name, and select the SystemPressure
command.
4. Enter the name of the pump currently delivering flow for this column.
5. For Chromeleon 6.8, click Execute.