3. Under TC, select the column name and select the FormatChipCard
command.
4. For Chromeleon 6.8, click Execute.
5. If the error occurs again, insert a new column ID chip card (see
Section
X position error.
This error occurs if a TC injection valve fails to switch position within the
time allotted.
To troubleshoot:
1. If a sequence is being executed, terminate the sequence by clicking Abort
Batch on the Sequence Control panel.
2. Press the
30 seconds, press the button again to turn on the power.
3. Try to toggle the injection valve from Load to Inject by clicking the Load
and Inject buttons on the TC ePanel or Control panel.
4. If the error occurs again, contact Technical Support for Dionex products
for assistance.
You do not have the privilege to approve the module's
qualification.
-or-
You do not have the privilege to change the performance limits.
-or-
You do not have the privilege to change the qualification
intervals.
In controlled environments, users are not allowed to perform certain
operations unless they have been granted the appropriate privileges. If you
attempt to do so, the appropriate message (shown above) is displayed.