The Warning threshold must be higher than the Limit threshold.
This error occurs if the Warning value selected for a Predictive Performance
command (for example, LeftRotorSealSwitches) is lower than the Limit
value. Chromeleon issues a warning message in the Audit Trail when the
Warning value is reached and an error message when the Limit value is
reached.
To troubleshoot:
Select a Warning value that is higher than the Limit value.
The Warning threshold must be lower than the Limit threshold.
This error occurs if the Warning value selected for a parameter is higher than
the Limit value. Chromeleon issues a warning message in the Audit Trail
when the Warning value is reached and an error message when the Limit
value is reached.
To troubleshoot:
Select a Warning value that is lower than the Limit value.
There was already an AcqOff command for this channel before.
-or-
There was already an AcqOn command for this channel before.
-or-
Two AcqOn commands for the same channel. The second
chromatogram will overwrite the first.
A Chromeleon 7 instrument method or Chromeleon 6.8 program can include
no more than one command to start or terminate data acquisition for a
channel; if multiple commands are present, the appropriate message (shown
above) appears.
To troubleshoot:
Remove the duplicate commands from the instrument method or program.