iCAP 6000 Series
Pre-Installation Manual
Version 4.2
Part number 8499 400 30001

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the iCAP 6000 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Thermo Scientific iCAP 6000 Series

  • Page 1 6000 Series Pre-Installation Manual Version 4.2 Part number 8499 400 30001 2007 Thermo Fisher Scientific. All rights reserved. Registration No. 441506 SOLAAR House 19 Mercers Row, Cambridge CB5 8BZ, United Kingdom. Telephone +44 (0) 1223 347400, Fax +44 (0) 1223 347402,...
  • Page 2 This page is intentionally blank Thermo Fisher Scientific iCAP 6000 Series Pre-Installation Manual v4.2 ii...
  • Page 3: Table Of Contents

    Water Temperature Control ............4-5 Waste Storage ................4-6 Spectrometer Fume Extraction .............4-6 Sample Fume Extraction...............4-8 Chapter 5 Communications Interface............5-9 On the Spectrometer ..............5-9 On the Data Station...............5-9 Chapter 6 Pre-Installation Checklist............6-11 Chapter 7 Special Agreements................7-1 Thermo Fisher Scientific iCAP 6000 Series Pre-Installation Manual v4.2...
  • Page 4 This page is intentionally blank. iCAP 6000 Series Pre-Installation Manual v4.2 Thermo Fisher Corporation...
  • Page 5: Introduction

    Chapter 1 Introduction This manual is designed to help ensure that your iCAP 6000 Series ICP-OES Spectrometer will be installed efficiently and will be able to meet your requirements quickly and completely. It will minimise installation time if the services and facilities detailed here are available before the equipment is delivered.
  • Page 6: System Accessories

    Station and iCAP 6000 accessory board) High Solids Sample Introduction kit HF Acid Sample Introduction kit Organics Sample Introduction Kit Volatile Organics Sample Introduction kit Please request site requirement guides for any purchased accessories iCAP 6000 Series Pre-Installation Manual v4.2 Thermo Fisher Scientific...
  • Page 7: Transportation

    920 mm 95 Kg (38.6 in) (35.4 in) (36.2 in) (210 Ib) Transportation to some destinations may necessitate additional packaging, for example a packing crate. Dimensions and exact specifications may vary. Thermo Fisher Scientific iCAP 6000 Series Pre-Installation Manual v 4.0...
  • Page 8: Short Term Storage

    Note: Warranty is not transferable to a third party without Thermo Fisher approval. iCAP 6000 Series Pre-Installation Manual v3.1 Thermo Fisher Scientific...
  • Page 9: Working Area Requirements

    Organic or volatile solvents should not be stored, even for a short time, near the instrument. Thermo Fisher Scientific iCAP 6000 Series Pre-Installation Manual v 4.2 4-1...
  • Page 10: Environmental Requirements

    Coding is blue (neutral), brown (live), green/yellow (earth/ground). Connection to the laboratory supply should be with an appropriately rated plug conforming to local electrical guide lines and local requirements. It must be possible to electrically isolate the instrumentation. iCAP 6000 Series Pre-Installation Manual v3.1 Thermo Fisher Corporation...
  • Page 11 Station PC, and for an autosampler, chiller, printer or any other additional accessory required. Each electrical outlet must have an effective earth/ground connection. This protection must not be negated by the use of an extension cable without a protective earth conductor. Thermo Fisher Corporation iCAP 6000 Series...
  • Page 12: Spectrometer Size

    85.5 Kg (32.7 in) (29.3 in) (23.2 in) (188 lbs) Table 4–2. Spectrometer Dimensions and Weight 831.2mm Figure 4–2. iCAP 6000 Series Front View Figure 4–3. iCAP 6000 Series Top View iCAP 6000 Series Pre-Installation Manual v3.1 Thermo Fisher Corporation...
  • Page 13: Gas Requirements

    Thermo Fisher organisation). Temperature variation must be less than 0.2°C per hour The flow requirement of the spectrometer is 5 l/min and a cooling capacity in excess of 900 W. Thermo Fisher Corporation iCAP 6000 Series...
  • Page 14: Waste Storage

    4-3 below. The extraction tube has an inlet diameter of 125 mm (5 in). iCAP 6000 Series Pre-Installation Manual v3.1 Thermo Fisher Corporation...
  • Page 15 Measure flow rate at instrument end of extraction system Figure 4–4. iCAP 6000 Extraction System PLEASE NOTE: EXTRACTION REQUIREMENTS ARE NOT THE SAME AS FOR PREVIOUS THERMO FISHER ICP SPECTROMETERS. Thermo Fisher Corporation iCAP 6000 Series...
  • Page 16: Sample Fume Extraction

    Working Area Requirements Sample Fume Additional, separate, extraction should be considered if significant numbers of volatile, or acidic, samples are left in the proximity of the Extraction instrumentation, even those in an autosampler. iCAP 6000 Series Pre-Installation Manual v3.1 Thermo Fisher Corporation...
  • Page 17: Communications Interface

    A separate Ethernet card will be required for networking A printer port maybe required A suitable PC is available from your local Thermo Fisher organisation. The installation engineer is not responsible for network and customer specific set-up of the PC. Thermo Fisher Corporation iCAP 6000 Series...
  • Page 18 Communications Interface This page is intentionally blank 5-10 iCAP 6000 Series Pre-Installation Manual v3.1 Thermo Fisher Corporation...
  • Page 19: Pre-Installation Checklist

    Maximum distance between the instrument and the regulator valve 99,995% 0 <10 Purity of argon (Gas or Liquid) ppm H 0<10 Pressure required at the input of spectrometer 6 bars (87 Psi) Thermo Fisher Corporation iCAP 6000 Series P...
  • Page 20 Optic nitrogen purge (Optional) Regulator valve in wall 0 – 10 bars (0 – 100 Psi) 6 mm (1/4 inch Gas union dimension at the exit of the regulator valve adapter supplied with instrument) Maximum distance between the instrument and the regulator valve 99,995% 0 <10...
  • Page 21: Special Agreements

    Chapter 7 Special Agreements This section should detail any special agreements, which have been arranged with your local Thermo Fisher organisation. Thermo Fisher Scientific iCAP 6000 Series Pre-Installation Manual v 4.2 7-1...