Analog Pressure Output
The analog pressure output indicates the operating pressure of the pump. The
pressure output is set to 50 mV/MPa (5 mV/14.51 psi). To monitor the
pressure, connect the pressure output to a recorder or an A/D converter. These
are the pin assignments for the 2-pin Cinch connector (P/N 8005.9001A):
Signal Level
Inner ring:
Outer ring:
32
Signal Name
Relay 2 Out
Gnd
Gnd
Gnd
Gnd
-----
Relay 4 Out
Relay 4 Out
Relay 4 Out
-----
Relay 3 Out
Relay 1 Out
Relay 2 Out
Vcc Out
TTL Input 1
TTL Input 2
TTL Input 3
-----
Table 2-1. 25-Pin D-Sub I/O Port (Female) (Continued)
Function
Signal (pressure)
Ground
Signal Level
Description
Potential-free
Common
Ground
Ground
Ground
Ground
Ground
Ground
Ground
Ground
-----
Not used
Potential-free
Normally open
Potential-free
Common
Potential-free
Normally closed
-----
Not used
Potential-free
Common
Potential-free
Normally open
Potential-free
Normally open
+5V
+5V, 500 mA
TTL
Hold/Run
TTL
Stop
TTL
Start
-----
Not used
Doc. 065342-06 12/11