Primary property "x" is already assigned and cannot be changed.
This error occurs if you attempt to change one of the primary column
properties (batch or lot number, date of manufacture, etc.) used for column
identification.
To troubleshoot:
Once a primary column property has been assigned, it cannot be changed. As
an alternative, consider changing a secondary column property. For a list of
primary and secondary properties, refer to the Chromeleon Help.
Properties cannot be changed. A chip card has not been inserted
properly. Check the CardState property for all cards and reinsert
those indicating "BusBlocked."
This error occurs if you attempt to change column ID properties when at least
one column ID chip card is inserted incorrectly.
To troubleshoot:
1. Open the Chromeleon 7 ePanel Set or the Chromeleon 6.8 panel tabset.
2. Press the F8 key to open the Commands dialog box.
3. Under TC, select the column name and select the CardState property. If
the status is BusBlocked, remove the card from the card slot and reinsert
it.
When the card is properly installed, the Dionex logo on the card faces up and
the LED on the memory slot is green.
Properties cannot be changed if the compartment door is open.
This error occurs if you attempt to change column ID properties when the TC
door is open (or is not fully closed).
To troubleshoot:
1. Verify that the door is fully closed.
2. If the door is not fully closed, check for—and remove—any obstruction.