Figure 2-52
voltammetry.
0.60
0.40
0.20
0.00
Volts
0.00
-0.20
-0.40
-0.60
(0.00, -0.80
-0.80
In this example, the potential is scanned from -0.80 to +0.60 V and then from
+0.60 to -0.80 V. The total time for one waveform period is 28 s. This
provides a sweep rate of 0.1 V/s as shown in the equation below.
106
shows an example of a triangular waveform used in cyclic
5.00
10.00
Figure 2-52. Example Cyclic Voltammetry Waveforms
2 V
----------------------------------- -
CV Cycle Time
2 1.4
--------------------
28s
(14.0, 0.60)
15.00
20.00
Seconds
=
Sweep Rate
V
V
-- -
=
0.1
s
25.00
30.00
(28.0, -0.80)
Doc. 065342-06 12/11