Notes for Capillary IC systems:
Capillary
IC
The 0.062 mm (0.0025 in) ID blue PEEK tubing (capillary tubing) used
for IC Cube connections is cut at the factory using a specialized precision
cutter. If you need to replace any of this type of tubing, order an IC Cube
Tubing Kit (P/N 072186) or the individual tubing listed in
When replacing capillary tubing, make sure the ferrule and fitting bolt are
at least 2 mm (0.1 in) from the end of the tubing before you insert the
tubing into the port. Do not position the ferrule and fitting bolt flush with
the end of the tubing. For detailed connection instructions, see
Except for capillary tubing, you can use a tubing cutter to cut tubing to the
required length. Make sure the cut is at a right angle to the length of the
tubing and there are no nicks or burrs on the end. A tubing cutter
(P/N 049584) is included in the DC Ship Kit (P/N 072011). Refer to the
instructions provided with the cutter.
For tightening requirements for the high-pressure fitting bolt
(P/N 074449) and ferrule (P/N 074373), see
For tightening requirements for the 10-32 fitting bolt (P/N 043275) and
ferrule (P/N 043276), refer to Installation of Dionex Liquid Line Fittings
(Document No. 031432). The manual is on the Thermo Scientific
Reference Library DVD (P/N 053891).
Doc. 065342-06 12/11