Thermo Scientific Dionex ICS-5000 Operator's Manual page 179

Ion chromatography system
Table of Contents

Advertisement

Linked Dual-Cartridge Configuration
In the linked dual-cartridge configuration, the EG contains two EGCs, linked
to a single pump. Note that the allowable eluent concentration for a linked
cartridge is less than when the cartridge is defined as independent. For details,
refer to the EGC manual. EGC manuals are included on the Thermo Scientific
Reference Library DVD (P/N 053891).
EGCs
K
CO
/EPM
2
3
Electrolytic pH
Modifier
KOH/KOH
KOH/MSA
KOH/NaOH
MSA/MSA
MSA/NaOH
NaOH/NaOH
LiOH/LiOH
KOH (Capillary)/
MSA (Capillary)
Doc. 065342-06 12/11
Eluent Concentration Range
0.1 to 15 mM at 0.1 to 1.0 ml/min
flow
0.1 to X mM at 1.0 to 2.0 ml/min
flow
where X = 15/flow
0.1 to 50 mM at 0.1 to1.0 mL/min
flow
0.1 to X mM at 1.0 to3.0 mL/min
flow
where X = 50/flow
0.1 to 40 mM at 0.1 to1.0 mL/min
flow
0.1 to X mM at 1.0 to3.0 mL/min
flow
where X = 40/flow
0.1 to 100 mM at 0.001 to
0.01 mL/min
0.1 to X mM at 0.01 to 0.1mL/min
where X = 1/flow
Table 4-3. Eluent Concentration Ranges for
Linked Dual-Cartridge Configurations
4 • EG Startup
Comment
The total of the eluent
concentrations from
both cartridges
(K
CO
and EPM)
2
3
must not exceed the
specified range. The
EPM concentration
must not exceed
10 mM. See the notes
below for additional
information.
The eluent
concentration range
for each cartridge is
50% of the range for
an independent
cartridge
The eluent
concentration range
for each cartridge is
50% of the range for
an single-independent
cartridge.
The eluent
concentration range
for each cartridge is
50% of the range for
an independent
cartridge.
163