3. Open the DC upper door.
4. Disconnect each liquid line connected to the IC Cube cartridge to be
replaced.
5. Loosen the two thumbscrews on the cartridge
(see
Figure
the IC Cube.
6. Slide the new cartridge into the IC Cube.
Push the cartridge firmly into the IC Cube
and then tighten the thumbscrews.
7. If you are installing a suppressor or CRD cartridge, hydrate them before
use. Refer to the instructions in the suppressor or CRD manual. The
manuals are included on the Thermo Scientific Reference Library DVD
(P/N 053891)
8. Reconnect the liquid lines that were disconnected in
Doc. 065342-06 12/11
9-40) and pull the cartridge out of
Figure 9-40. Capillary Cartridge Thumbscrews
9 • DC Service
(EG Degas Cartridge Shown)
Step
4.
Cartridge
Thumb-
screws
373