Dionex ICS-5000 Operator's Manual
Chromeleon provides several pre-programmed waveforms.
You can use a pre-programmed waveform, without modification, or modify it
for your application. You can also define a new waveform. Waveforms are
defined and modified in the Waveform Editor (see
To view the Waveform Editor, click Edit on the ED Options page in the
Instrument Method Wizard (Chromeleon 7) or Program Wizard (Chromeleon
6.8).
Analytical vs. Scanning Waveforms
The pre-programmed analytical waveforms supplied with Chromeleon are
designed for quantitative analysis of specific compounds (alcohols, amino
acids, carbohydrates, etc.). With analytical waveforms, integration occurs
either while a single constant potential is being applied over time (see the
example waveform in
being applied (see the example waveform in
108
Figure 2-54. Waveform Editor in Chromeleon 7
Figure
2-50) or while a series of stepped potentials is
Figure
2-54).
Figure
2-51).