Current linked pump is capillary. Please install a capillary EGC.
-or-
Current linked pump is analytical. Please install an analytical
EGC.
This error occurs if the type of pump linked to the EGC does not match the
EGC type. A capillary EGC can only be used with a capillary pump and an
analytical EGC can only be used with an analytical pump.
To troubleshoot:
1. Install the correct type of EGC for the pump type.
2. For a hybrid system (a dual system with both an analytical and a capillary
pump), check the EGC configuration properties in the Chromeleon 7
Instrument Controller Service or the Chromeleon 6.8 Server
Configuration program to verify that the correct pump is linked to the
EGC.
EG1 cartridge disconnected.
-or-
EG2 cartridge disconnected.
This error occurs if Chromeleon sends a command to set an EG parameter
when the EGC is disconnected.
To troubleshoot:
1. Make sure the EGC cable is securely plugged into the EG (see
Figure
2. If the error message appears again, contact Technical Support for Dionex
products for assistance. The EGC control electronics may have
malfunctioned.
NOTE
The Thermo Scientific Dionex ICS-5000 Ion Chromatogra-
phy System electronics components cannot be serviced by the
user.
244
2-10) and the locking ring is tightened.
Doc. 065342-06 12/11