Dionex ICS-5000 Operator's Manual
4. Disconnect each liquid line connected to the valve.
5. Unscrew the locking ring on the outside of the valve (see
and remove the ring.
Figure 9-35. Unscrew the Locking Ring (Valve in IC Cube Shown)
6. Grasp the front of the valve pod and pull out firmly to remove it from the
DC or AM.
7. Align the slots in the new pod with the runner in the valve holder on the
DC or AM (see
(one slot is narrower than the other). Verify that the slots are aligned with
their matching runners.
368
Figure
9-36). Valve pods are keyed to fit only one way
Figure
9-35)
Locking Ring
Doc. 065342-06 12/11