Thermo Scientific Dionex ICS-5000 Operator's Manual page 381

Ion chromatography system
Table of Contents

Advertisement

Um Verletzungen zu vermeiden, warten Sie bitte, bis der DC abgekühlt
ist, ehe Sie die Tür öffnen bzw. Wartungs oder Servicearbeiten
durchführen.
3. Open the DC lower door.
4. Loosen the screw on the front of the leak sensor (P/N 062437) (see
Figure
5. Pull out the leak sensor from the
component panel and continue
pulling until the attached cable
is also out of the opening.
6. The cable attached to the leak
sensor connects to a cable
inside the DC. Continue pulling
the cable until the connectors
for the two cables are outside
the panel (see
7. Disconnect the two cables.
8. Connect the cable from the new leak sensor to the DC cable.
9. Feed the cables back inside the DC. Align the leak sensor with the
component panel opening and fingertighten the screw.
10. Make sure the leak sensor does not touch the bottom of the drip tray.
Doc. 065342-06 12/11
9-33). Note: The screw remains attached to the sensor.
Figure 9-33. Leak Sensor
Figure
9-34).
Leak
Sensor
Loosen
this
screw
Figure 9-34. Leak Sensor Cable
9 • DC Service
Disconnect
cables
365